One For All HFC-302 Bedienungsanleitung Seite 1

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Messen, Testen & Steuern One For All HFC-302 herunter. One For ALL HFC-302 User's Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - HASTINGS

INSTRUCTION MANUAL HFM-300 FLOW METER, HFC-302 FLOW CONTROLLER TELEDYNE HASTINGS INSTRUMENTS

Seite 2 - Manual Print History

300-302 Series Page 10 of 31 Span Error vs Pressure(0.026" Sensor Tube)y = 9.8877E-11x33.4154E-07x - 28.3288E-05x + -20.0%-18.0%-16.0%-14.0%-1

Seite 3 - Table of Contents

300-302 Series Page 11 of 31 Span Error Vs. Pressure0.017" Sensor-2%-1%0%1%2%3%4%5%0 100 200 300 400 500 600 700 800 900 1000Pressure (psig)Sp

Seite 4 - 1. General Information

300-302 Series Page 12 of 31 External Input pin (See Tables 2.1 & 2.2), the ratio of flows can be maintained over the entire range of gas flows

Seite 5 - 1.2. Specifications

300-302 Series Page 13 of 31 On the controller card there is a jumper that sets whether the control loop controls mass flow or an external process

Seite 6 - 1.4. Other Accessories

300-302 Series Page 14 of 31 Fig. 2.8 2.12. Temperature Coefficients As the ambient temperature of the instrument changes from the original calibr

Seite 7 - Quick Start

300-302 Series Page 15 of 31 3. Theory of Operation This section contains an overall functional description of the Hastings 300 series of flow ins

Seite 8 - 2.5. Electrical Connections

300-302 Series Page 16 of 31 can be dissipated by radiation, conduction, or convection. The radiation term is negligible due to the low temperature

Seite 9 - Fig. 2.1 Fig. 2.2

300-302 Series Page 17 of 31 heater to the gas stream will force the upstream bridge control loop to apply more power to the upstream heater so tha

Seite 10 - Mean error

300-302 Series Page 18 of 31 desired to ensure a stable instrument, there must be some matching between the linear volumetric flow versus pressure

Seite 11 - SensorPPPError

300-302 Series Page 19 of 31 πμρ5QLe= For a typical flow divider tube the entry length is approximately 0.16 cm. From t

Seite 12 - Fig. 2.6

300-302 Series Page 2 of 31 Manual Print History The print history shown below lists the printing dates of all revisions and addenda created for th

Seite 13 - Fig. 2.7

300-302 Series Page 20 of 31 00.020.040.060.080.10.120.140.160.180 5 10 15 20 25 30 35Flow (liter/min)Passage Thickness (cm)Fig. 3.5 Thickness of t

Seite 14 - Fig. 2.8

300-302 Series Page 21 of 31 3.7. Control Valve The control valve is an “automatic metering solenoid” valve (see Figure 3.7). While most solenoi

Seite 15 - 3. Theory of Operation

300-302 Series Page 22 of 31 This section contains service and calibration information. Some portions of the instrument are delicate. Use extrem

Seite 16 - TTCmP −=−

300-302 Series Page 23 of 31 Action: Shut off all flow. For the standard 0-5VDC output, adjust the zero potentiometer located on the upper r

Seite 17 - Fig. 3.2

300-302 Series Page 24 of 31 reverse order of disassembly. A new nickel gasket will be required. Secure the endcap with 65 in lb. (7.3 N m) to 85

Seite 18 - Fig. 3.3

300-302 Series Page 25 of 31 5. Gas Conversion Factors Gas conversion factors (GCF’s) for gasses metered using Hastings Instruments products, can

Seite 19

300-302 Series Page 26 of 31 being used to measure the flow of propane. The reading from the meter is multiplied by the GCF for propane divided by

Seite 20

300-302 Series Page 27 of 31 6. Volumetric Vs Mass Flow Mass flow measures just what it says, the mass or number of molecules of the gas flowing t

Seite 21 - Fig. 3.7

300-302 Series Page 28 of 31 7. Drawings and References HFM-300 / HFC-302 Outline Drawing (All dim. Are approximate)

Seite 22 - 4. Maintenance

300-302 Series Page 29 of 31 TRANSDUCER ENDCAP ENDCAP SHUNT HFM-300 Sectional View Sensor and Main PC Board

Seite 23 - Fig. 4.1

300-302 Series Page 3 of 31 Table of Contents 1. GENERAL INFORMATION...

Seite 24 - 4.5. Sensor Replacement

300-302 Series Page 30 of 31 SHUNT ENDCAP ENDCAPVALVE ASSEMBLY CONTROLLER CARD TRANSDUCER HFC-302 Sectional View

Seite 25 - 5. Gas Conversion Factors

300-302 Series Page 31 of 31 8. WARRANTY 8.1. Warranty Repair Policy Hastings Instruments warrants this product for a period of one year from the

Seite 26

300-302 Series Page 4 of 31 1. General Information The Teledyne Hastings HFM-300 is used to measure mass flow rates in gases. In addition to f

Seite 27 - 6. Volumetric Vs Mass Flow

300-302 Series Page 5 of 31 1.2. Specifications Accuracy ... <

Seite 28 - 7. Drawings and References

300-302 Series Page 6 of 31 The 4-20 mA I/O option can accept a current input. The 0-5 VDC command signal on pin 14 can be replaced by a 4-20mA co

Seite 29

300-302 Series Page 7 of 31 2. Installation and Operation This section contains the steps necessary to assist in getting a new flow meter/control

Seite 30

300-302 Series Page 8 of 31 2.4. Mechanical Connections 2.4.1. Filtering The smallest of the internal passageways in the Hastings 300 is the diame

Seite 31 - 8. WARRANTY

300-302 Series Page 9 of 31 Figures 2.1/2.2, and Tables 2.1/2.2, show the 300/302 pin out. Pin # Pin #1 Signal Common 1 Do not use2 D

Kommentare zu diesen Handbüchern

Keine Kommentare